首页 /研究 /Development of a MR-compatible cable-driven manipulator: Design and technological issues
MANIPULATION

Development of a MR-compatible cable-driven manipulator: Design and technological issues

Salih Abdelaziz, Laure Esteveny, Laurent Barbé, Pierre Renaud, Bernard Bayle, Michel de Mathelin

发表年份
2012
引用次数
14

摘要

In this paper, we focus on the technology issues to be solved to develop a cable-driven robot compatible with Magnetic Resonance Imaging constraints. This study is based on the design of a new compact cable-driven manipulator with remote actuators, initially developed for prostate interventions. One of the originalities of the system is to use an instrumented structure to evaluate the cable tensions and lengths in order to perform an adequate control. The sensors assessment has been experimentally achieved and the necessity to introduce a new control strategy using the developed sensors has been demonstrated.

关键词

ActuatorRobot manipulatorManipulator (device)Focus (optics)Computer scienceRobotRemote controlEngineeringControl engineeringElectrical engineering