A Comparative Experimental Study Evaluating the Performance of Surgical Robots Aesop and Endosista
Yunus Yavuz, Brynjulf Ystgaard, Eirik Skogvoll, Ronald M rvik
- 发表年份
- 2000
- 引用次数
- 38
摘要
The aim of this study was to experimentally assess and compare the accuracy of the surgical robots Aesop and Endosista as camera holders for use in laparoscopic surgery. The performance of these two robotic systems was examined for linear (upwards, downwards, diagonal), complex, and "in and out" movements using laparoscopic training boxes. Standard distances and tests were used for each system, and the time required to achieve each task was measured. The majority of the linear movements of the verbal and preprogrammed modes of Aesop were quicker than those of Endosista. Diagonal movements were significantly faster with the preprogrammed-mode Aesop. Complex or three-dimensional movements were also significantly faster with both modes of Aesop than with Endosista. Under the experimental conditions, Aesop, particularly in the preprogrammed mode, is quicker and more accurate than Endosista.
关键词
相关论文
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
A new optimizer using particle swarm theory
R.C. Eberhart, James Kennedy
2002