A study of needle image artifact localization in confirmation imaging of MRI-guided robotic prostate biopsy

Sang‐Eun Song, Nathan B. Cho, Iulian Iordachita, Peter Guion, Gábor Fichtinger, Louis L. Whitcomb

发表年份
2011
引用次数
10

摘要

Recently several systems for magnetic resonance image (MRI) guided needle placement in the prostate have been reported. In comparison to conventional ultrasound-guided needle placement in the prostate, these MRI-guided systems promise improved targeting accuracy for prostate intervention procedures including biopsy, fiducial marker insertion, injection and focal therapy. In MRI-guided needle interventions, after a needle is inserted, the needle position is often confirmed with a volumetric MRI scan. Commonly used titanium needles are not directly visible in an MR image, but they generate a susceptibility artifact in the immediate neighborhood of the needle. This paper reports the results of a quantitative study of the relation between the true position of titanium biopsy needle and the corresponding needle artifact position in MR images. The titanium needle artifact was found to be displaced 0.38 mm and 0.32 mm shift in scanner's frequency and phase encoding direction, respectively. The artifact at the tip of the titanium needle was observed to bend toward the scanner's B(0) magnetic field direction.

关键词

Artifact (error)Fiducial markerMagnetic resonance imagingProstateScannerMedicineRadiologyBiopsyUltrasoundNuclear medicine