Biopsy Needle Artifact Localization in MRI-Guided Robotic Transrectal Prostate Intervention

Sang‐Eun Song, Niasaki Cho, Iulian Iordachita, Peter Guion, Gábor Fichtinger, A. Kaushal, Kevin Camphausen, Louis L. Whitcomb

发表年份
2012
引用次数
18

摘要

Recently a number of robotic intervention systems for magnetic resonance image (MRI)-guided needle placement in the prostate have been reported. In MRI-guided needle interventions, after a needle is inserted, the needle position is often confirmed with a volumetric MRI scan. Commonly used titanium needles are not directly visible in an MRI, but they generate a susceptibility artifact in the immediate neighborhood of the needle. This paper reports the results of a quantitative study of the relationship between the true position of titanium biopsy needle and the corresponding needle artifact position in MRI, thereby providing a better understanding of the influence of needle artifact on targeting errors. The titanium needle tip artifact extended 9 mm beyond the actual needle tip location with tendency to bend toward the scanner's B (0) magnetic field direction, and axially displaced 0.38 and 0.32 mm (mean) in scanner's frequency and phase encoding direction, respectively.