Robotic System for Closed-Bore MRI-Guided Prostatic Interventions
A.A. Goldenberg, John Trachtenberg, Walter Kucharczyk, Yang Yi, Masoom A. Haider, Leo Ma, Robert Weersink, Cyrus Raoufi
- 发表年份
- 2008
- 引用次数
- 69
摘要
This paper reports on the development of a new closed-bore magnetic resonance imaging (MRI) compatible robotic system for image-guided prostatic interventions: ablation, brachytherapy, and biopsy. The first stage of development addresses only laser-based ablation. The robot actuators are ultrasonic motors. The first physical robot prototype was manufactured and tested in the MRI with an ablation tool. The tests covered magnetic resonance (MR) compatibility, tool visualization, and robot control accuracy. Robot tip position error is less than 2 mm at points closer than 0.5 m to the isocenter. A method to control ultrasonic motors for MRI-compatibility is reported.