Force Feedback Control System Dedicated for Robin Heart Surgical Robot

Zbigniew Nawrat, Kamil Rohr, Péter Fürjes, Łukasz Mucha, Krzysztof Lis, J Radó, Csaba Dücső, Péter Földesy, W Sadowski, Dariusz Krawczyk, Piotr Kroczek, Gábor Szebényi, Pál Soós, Z. Małota

发表年份
2016
引用次数
9

摘要

3D contact force sensors were developed and integrated in a demonstration system for testing the feasibility of their application in minimal invasive surgery (MIS). Piezoresistive MEMS based vectorial force sensors were designed and fabricated by 3D silicon micromachining technology and packaged according to their proposed transducer and sensor applications. In this work we demonstrated the integrability and functional applicability of the 3D force sensors in MIS robotic systems to improve their flexibility and reliability by providing real-time force and tactile information during the operation. The final goal is to integrate the new subsystems in the Robin Heart surgery robot of FRK. [1] Three different functions are targeted: 1. Micro-joystick actuator to be integrated in the hilt of the laparoscope to easily control robotic movement during operation. 2. Force sensor inside the laparoscopic jaw to provide feedback to the surgeon by measuring the grasping strength and 3. 3D force/tactile sensor which facilitates palpation for tissue diagnostics during operation. In this paper we demonstrate a feasibility study regarding these proposed applications.

关键词

JoystickTactile sensorTeleoperationPiezoresistive effectActuatorHaptic technologyFlexibility (engineering)RobotContact forceEngineering