首页 /研究 /Constrained motion control of multisegment continuum robots for transurethral bladder resection and surveillance
SURGICAL

Constrained motion control of multisegment continuum robots for transurethral bladder resection and surveillance

Andrea Bajo, Ryan B. Pickens, S. Duke Herrell, Nabil Simaan

发表年份
2013
引用次数
18

摘要

Constrained motion control of robotic end-effectors is essential for safe operation in confined spaces such as the urinary bladder. This paper presents the clinical motivation for the development of new control algorithms for robotic-assisted transurethral bladder resection and surveillance using multisegment continuum robots. The anatomy, workspace, and access constraints for this procedure are identified and used as a guideline for the design of the telesurgical system and its control architecture. Constraints are mapped into the configuration space of the robot rather than in task space simplifying the modeling and the enforcement of virtual fixtures. The redundancy resolution is autonomously modified in order to exploit the remaining degrees of freedom using task priority. These methods are validated on a glass model of urinary bladder.

关键词

WorkspaceRobotRedundancy (engineering)Robot end effectorComputer scienceMotion controlArtificial intelligenceSimulationComputer vision

相关论文

查看 SURGICAL 分类全部论文