Home /Research /Cardiac interventions under MRI guidance using robotic assistance
SURGICAL

Cardiac interventions under MRI guidance using robotic assistance

Ming Li, Ankur Kapoor, Dumitru Mazilu, Bradford J. Wood

Year
2010
Citations
9

Abstract

Transapical aortic valve replacement under MRI guidance in a beating heart is a recent minimally invasive technique that could benefit from a surgical assistant system. We present a robotic surgical assistant system that can precisely and repeatably deliver aortic valve prostheses. The surgical system consists of an imaging system, an Innomotion robotic arm, a 3-DoF valve delivery module and user interfaces. Interactive control allows the physician to remain in the loop and adjust the orientation and position using real-time MR feedback. The 3-DoF valve delivery module is developed to deploy both balloon-expandable and self-expanding stented prostheses. We use a new compact fiducial that can be placed close to the volume of interest and requires a single image plane for image based robot registration. We evaluate the MRI compatible valve delivery module for both types of prostheses. The accuracy for prosthesis delivery is about 0.8 mm and 1.5 mm, for self-expanding and balloon-expandable prostheses, respectively. Preliminary results in ex-vivo experimentation suggest that the robotic system can be translated into animal and clinical models.

Keywords

Fiducial markerSurgical robotOrientation (vector space)Biomedical engineeringComputer scienceGuidance systemProsthesisRobotMechanical valveComputer vision