Toward Intra-Operative Prostate Photoacoustic Imaging: Configuration Evaluation and Implementation Using the da Vinci Research Kit
Hamid Moradi, Shuo Tang, Septimiu E. Salcudean
- Year
- 2018
- Citations
- 33
Abstract
We compare different possible scanning geometries for prostate photoacoustic tomography (PAT) while considering a realistic reconstruction scenario in which the limited view of the prostate and the directivity effect of the transducer are considered. Simulations and experiments confirm that an intra-operative configuration in which the photoacoustic signal is received by a pickup transducer from the anterior surface of the prostate provides the best approach. We propose a PAT acquisition system that includes a da Vinci system controlled by the da Vinci Research Kit, an illumination laser, and an ultrasound machine with parallel data acquisition. The robot maneuvers the pickup transducer to form a cylindrical detection surface around the prostate. The robot is programmed to acquire trajectories in which the transducer face is parallel to and oriented toward a rotational tomography axis, while the laser is fired and PAT data are collected at regular intervals. We present our initial images acquired with this novel system.
Keywords
Related papers
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Fractional Differential Equations
Igor Podlubný
2025
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991