MR-Conditional Robotic Actuation of Concentric Tendon-Driven Cardiac Catheters
Yifan Wang, Zheng Qiu, Junichi Tokuda, Ehud J. Schmidt, Aravindan Kolandaivelu, Yue Chen
- 发表年份
- 2024
- 引用次数
- 3
摘要
Atrial fibrillation (AF) and ventricular tachycar-dia (VT) are two of the sustained arrhythmias that significantly affect the quality of life of patients. Treatment of AF and VT often requires radiofrequency ablation of heart tissues using an ablation catheter. Recent progress in ablation therapy leverages magnetic resonance imaging (MRI) for higher contrast visual feedback, and additionally utilizes a guiding sheath with an actively deflectable tip to improve the dexterity of catheter ma-nipulation inside the heart. This paper presents the design and validation of an MR-conditional robotic module for automated 6-degrees-of-freedom actuation of both the ablation catheter and the sheath. The robotic module features a compact design for improved accessibility inside the MR scanner bore and is driven by piezoelectric motors to ensure MR-conditionality. The combined catheter-sheath mechanism is essentially a concentric tendon-driven continuum robot and its kinematics is modeled by the constant curvature model for closed-loop position control. Path following experiments were conducted to validate the actuation module and control scheme, achieving < 2 mm average tip position error with closed-loop control and online shape fitting.