A coordinate positioning puncture method under robot-assisted CT-guidance: phantom and animal experiments

Weifan Zhang, Peng Xia, Shijie Liu, Xiaowei Huang, Xinhui Zhao, Zhao Liu, Hui Dang, Xiaohu Li, Gang Niu

发表年份
2020
引用次数
11

摘要

Purpose To evaluate the accuracy of the robot-assisted computed tomography (CT)-guided coordinate positioning puncture method by phantom and animal experiments.Material and methods In the phantom experiment, seven robot-assisted punctures were made to evaluate the accuracy of the method. In the animal experiment, 18 punctures (nine robotic and nine manual) were made in the livers of nine rabbits. The indicators, such as needle-tract length, angle deviation, puncture accuracy, number of scans required, and radiation exposure dose were compared between manual and robotic punctures. The paired-samples t-test was used for analysis.Results In the phantom experiment, the mean accuracy of seven punctures was 2.67 mm. In the animal experiment, there was no significant difference in needle-tract length (32.58 mm vs. 34.04 mm, p = .606), angle deviation (17.21° vs. 21.23° p = .557) and puncture accuracy (8.42 vs. 8.77 mm, p = .851) between the two groups. However, the number CT scans required (2.44 vs. 3.33, p = .002), and the radiation exposure dose (772.98 vs. 1077.89 mGy/cm, p = .003) were lower in the robot group than in the manual group.Conclusions The coordinate positioning puncture method under robot-assisted CT-guidance can reach an accuracy that is comparable to that of the traditional manual CT-guided puncture method and with fewer CT scanning times accompanied with a lower radiation dosage.

关键词

Imaging phantomNuclear medicineMedicineBiomedical engineeringAbsolute deviationComputed tomographySignificant differenceRadiologyMathematics

相关论文

查看 OTHER 分类全部论文