Large area kidney imaging for pre-transplant evaluation using real-time robotic optical coherence tomography
Xihan Ma, Mousa Moradi, Xiaoyu Ma, Qinggong Tang, Moshe Levi, Yu Chen, Haichong K. Zhang
- 发表年份
- 2024
- 引用次数
- 17
- 访问权限
- 开放获取
摘要
Optical coherence tomography (OCT) can be used to image microstructures of human kidneys. However, current OCT probes exhibit inadequate field-of-view, leading to potentially biased kidney assessment. Here we present a robotic OCT system where the probe is integrated to a robot manipulator, enabling wider area (covers an area of 106.39 mm by 37.70 mm) spatially-resolved imaging. Our system comprehensively scans the kidney surface at the optimal altitude with preoperative path planning and OCT image-based feedback control scheme. It further parameterizes and visualizes microstructures of large area. We verified the system positioning accuracy on a phantom as 0.0762 ± 0.0727 mm and showed the clinical feasibility by scanning ex vivo kidneys. The parameterization reveals vasculatures beneath the kidney surface. Quantification on the proximal convoluted tubule of a human kidney yields clinical-relevant information. The system promises to assess kidney viability for transplantation after collecting a vast amount of whole-organ parameterization and patient outcomes data.