Toward new minimally invasive surgical robotic system
Med Amine Laribi, Marc Arsicault, Thomas Riviere, Saïd Zeghloul
- 发表年份
- 2012
- 引用次数
- 18
摘要
This paper present a new laparoscopic surgical setup (minimally invasive surgery, MIS) with spherical serial surgical robot. The robot was developed at the PPRIME Institute in the ROBIOSS team from Poitiers University. Based on laparoscopic surgery, specification for workspace and constraints are outlined and the architecture selected for the surgical robots is presented and discussed. Analytical models of the inverse and forward kinematics are provided together with a detailed analysis of mobility and constraints. The proposed approach is based on motion capture of an expert gestures during which an anastomosis technique has to be performed. The evaluation of the workspace is based on the use of the Vicon Nexus motion capture system. The medical gestures were analyzed in term of position and velocities; the result has been used in the definition of the kinematics specifications of the proposed manipulator. The effective workspace size is determined through an experimental study on a simulator. The spherical serial robot (3R-T) has been selected because of its characteristics meeting the constraint requirements. The robot and its environment are described.
关键词
相关论文
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
A new optimizer using particle swarm theory
R.C. Eberhart, James Kennedy