首页 /研究 /A comprehensive system for dosimetric commissioning and Monte Carlo validation for the small animal radiation research platform
OTHER

A comprehensive system for dosimetric commissioning and Monte Carlo validation for the small animal radiation research platform

E. Tryggestad, Michael Armour, Iulian Iordachita, Frank Verhaegen, John Wong

发表年份
2009
引用次数
81

摘要

Our group has constructed the small animal radiation research platform (SARRP) for delivering focal, kilo-voltage radiation to targets in small animals under robotic control using cone-beam CT guidance. The present work was undertaken to support the SARRP's treatment planning capabilities. We have devised a comprehensive system for characterizing the radiation dosimetry in water for the SARRP and have developed a Monte Carlo dose engine with the intent of reproducing these measured results. We find that the SARRP provides sufficient therapeutic dose rates ranging from 102 to 228 cGy min(-1) at 1 cm depth for the available set of high-precision beams ranging from 0.5 to 5 mm in size. In terms of depth-dose, the mean of the absolute percentage differences between the Monte Carlo calculations and measurement is 3.4% over the full range of sampled depths spanning 0.5-7.2 cm for the 3 and 5 mm beams. The measured and computed profiles for these beams agree well overall; of note, good agreement is observed in the profile tails. Especially for the smallest 0.5 and 1 mm beams, including a more realistic description of the effective x-ray source into the Monte Carlo model may be important.

关键词

Monte Carlo methodRangingDosimetryRadiationBeam (structure)Range (aeronautics)PhysicsComputational physicsOpticsNuclear medicine

相关论文