Robotic laser systems offer the potential for sub-millimeter, non-contact, high-precision tissue resection, yet existing platforms lack volumetric planning and intraoperative feedback. We present RATS (Robot-Assisted Tissue Surgery), an intelligent opto-mechanical, optical coherence tomography (OCT)-guided robotic platform designed for autonomous volumetric soft tissue resection in surgical applications. RATS integrates macro-scale RGB-D imaging, micro-scale OCT, and a fiber-coupled surgical laser, calibrated through a novel multistage alignment pipeline that achieves OCT-to-laser calibration accuracy of 0.161+-0.031mm on tissue phantoms and ex vivo porcine tissue. A super-Gaussian laser-tissue interaction (LTI) model characterizes ablation crater morphology with an average RMSE of 0.231+-0.121mm, outperforming Gaussian baselines. A sampling-based model predictive control (MPC) framework operates directly on OCT voxel data to generate constraint-aware resection trajectories with closed-loop feedback, achieving 0.842mm RMSE and improving intersection-over-union agreement by 64.8% compared to feedforward execution. With OCT, RATS detects subsurface structures and modifies the planner's objective to preserve them, demonstrating clinical feasibility.