Matthew J. Gerber, Jean‐Pierre Hubschman, Tsu‐Chin Tsao

Year
2021
Citations
10

Abstract

BACKGROUND: In cataract surgery, polishing of the posterior capsule (PC) can lead to improved surgical outcomes but is currently avoided due to its high-risk nature. This work developed a robotic system capable of performing PC polishing on ex vivo pig eyes using optical coherence tomography (OCT) guidance. METHODS: The lenses of five ex vivo pig eyes were extracted and a thin layer of glue deposited onto the PC. Transpupillary OCT scans of the anterior segment were used to generate a PC-polishing trajectory. During polishing, OCT B-scans tracked the tool tip and were displayed to the operator. RESULTS: Complete removal of the glue was accomplished in all five trials with no PC rupture reported. CONCLUSIONS: The feasibility of using a robotic system guided by OCT to perform PC polishing on a biological model was demonstrated. Contributions include modelling of the PC anatomy, intraoperative OCT visualization, and automated tool-tip motion with scheduled aspiration pressures.