Laparoscopic Camera Based on an Orthogonal Magnet Arrangement
Nicolò Garbin, Piotr R. Slawinski, Gregorio Aiello, Christina Karraz, Pietro Valdastri
- Year
- 2016
- Citations
- 22
Abstract
In this letter, we present for the first time a magnetic anchoring-actuation link with an auto-flip feature. This orthogonal magnetic arrangement relies on the placement of two permanent magnets such that their magnetic moments are respectfully orthogonal. Though the arrangement may have many applications, in this study we integrate it in a small factor magnetic camera for minimally invasive procedures. Upon insertion through a trocar incision, the 5.5 mm diameter and 35 mm length magnetic camera is coupled with an external robotic controller and displaced from the port thus preventing clutter of the surgical workspace. The device allows for manual lateral translation as well as robotically controlled tilt and pan, resulting in four degrees of freedom. The auto-flip feature prevents the need for image adjustment in software as the camera tilts through its hemispherical workspace. A static model that relates an input external control tilt and output camera tilt has been developed and validated. Favorable results during bench and canine cadaver evaluation suggest promise for the proposed magnetic camera to improve the state of art in minimally invasive surgical procedures.