Home /Research /Robot‐guided stereoelectroencephalography without a computed tomography scan for referencing: Analysis of accuracy
SURGICAL

Robot‐guided stereoelectroencephalography without a computed tomography scan for referencing: Analysis of accuracy

Andrea Spyrantis, Adriano Cattani, Adam Strzelczyk, Felix Rosenow, Volker Seifert, Thomas M. Freiman

Year
2018
Citations
31

Abstract

OBJECTIVE: Recent studies with robot-guided stereotaxy use computed tomography (CT) scans for referencing. We will provide evidence that using preoperative MRI datasets referenced with a laser scan of the patient's face is sufficient for sEEG implantation. METHODS: In total, 40 sEEG electrodes were implanted in five patients by the robotic surgical assistant (ROSA). The postoperative CT scan for identifying electrode positions was fused with the preoperative MRI-based planning data. The accuracy was determined by the target point error (TPE) and the entry point error (EPE), applying the Euclidean distance. RESULTS: The mean TPE amounted to 2.96 mm, the mean EPE to 2.53 mm. The accuracy was improved in 1.5 T MRI: the mean TPE amounted to 1.72 mm, the EPE to 0.97 mm. No complications, haemorrhages, infections, etc., were observed. CONCLUSIONS: Robot-guided sEEG based on 3 T MRI reduces radiation exposure for the patient and can still be performed safely.

Keywords

StereoelectroencephalographyStereotaxyNuclear medicineMedicineComputed tomographyRadiologyComputer scienceArtificial intelligenceElectroencephalographyHaptic technology

Related papers

Browse all SURGICAL papers