Modeling and Control of a 2-DoF Meso-Scale Continuum Robotic Tool for Pediatric Neurosurgery
Yash Chitalia, Seokhwan Jeong, Kent K. Yamamoto, Joshua J. Chern, Jaydev P. Desai
- Year
- 2020
- Citations
- 74
Abstract
This article introduces the analysis and control of a meso-scale two degree-of-freedom robotic endoscopic tool body for minimally invasive surgeries. The design of the robotic tool uses two types of a tendon-driven joint known as a bending flexure joint that allows us to control each degree-of-freedom by minimizing interjoint coupling by design. Pure kinematic modeling and control for these robots may not provide precise control performance due to kinematic uncertainties arising from tendon elongation, tendon slacking, gear backlash, etc. We propose a static model for each of the joints of the robotic tool that avoids several of these problems. Depending on the direction of tendon tension application, the proximal joint displays considerable hysteresis due to the superelastic material characteristics and this is included in our static model. The statics of a highly compliant distal joint is also modeled and validated using finite element analysis and experimental data. Using these models, we develop a control system that comprises of a disturbance observer and the proposed static model to provide precise force control and compensate for joint hysteresis.
Keywords
Related papers
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
A new optimizer using particle swarm theory
R.C. Eberhart, James Kennedy
2002