Electromagnetic braking-based collision protection of a novel catheter manipulator
Linshuai Zhang, Shuxiang Guo, Huadong Yu, Shuoxin Gu, Yu Song, Miao Yu
- Year
- 2017
- Citations
- 11
Abstract
A robotically assisted catheterization system can obviously reduce the radiation exposure to the surgeon and lesson the fatigue caused by standing for long time in protective clothing. However, few designs have taken the collision protection function of a catheter manipulator into consideration. Additionally, limited research has been conducted in the damage of the clamping mechanism to the catheter. This paper presents a novel clamping mechanism based on the electromagnetic braking for a catheter manipulator which can be used in the minimally invasive surgery training system. A significant advantage is that the proposed design has the collision protection function which can realize the relative sliding between the catheter and the clamping device so as to avoid the vessel puncture, when the measured force exceeds a certain threshold. In addition, the simulation verification of this function is carried out. The results show that the clamping force of the catheter obviously decreases when the measured force increases. This clamping mechanism provides important insights into the design of safe and reliable robotic catheter manipulators incorporating effective and lossless clamping for intraoperative navigation.