A Clinical Tuning Framework for Continuous Kinematic and Impedance Control of a Powered Knee-Ankle Prosthesis

Emma Reznick, T. Kevin Best, Robert D. Gregg

发表年份
2025
引用次数
1

摘要

Objective: Configuring a prosthetic leg is an integral part of the fitting process, but the personalization of a multi-modal powered knee-ankle prosthesis is often too complex to realize in a clinical environment. This paper develops both the technical means to individualize a hybrid kinematic-impedance controller for variable-incline walking and sit-stand transitions, and an intuitive Clinical Tuning Interface (CTI) that allows prosthetists to directly modify the controller behavior. Methods and procedures: Utilizing an established method for predicting kinematic gait individuality alongside a new parallel approach for kinetic individuality, we personalize continuous-phase/task models of joint impedance (during stance) and kinematics (during swing) using tuned characteristics exclusively from level-ground walking. To take advantage of this method, we developed a CTI that translates common clinical tuning parameters into model adjustments for the walking and sit-stand controllers. We then conducted a case study where a prosthetist iteratively tuned the powered prosthesis to an above-knee amputee participant in a simulated clinical session involving sit-stand transitions and level walking, from which incline/decline walking features were automatically calibrated. Results: The prosthetist fully tuned the multi-activity prosthesis controller in under 20 min. Each iteration of tuning (i.e., observation, parameter adjustment, and model reprocessing) took 2 min on average for walking and 1 min on average for sit-stand. The tuned behavior changes were appropriately manifested in the commanded prosthesis torques, both at the manually tuned tasks and automatically tuned tasks (inclines). Conclusion: The CTI leveraged able-bodied trends to efficiently personalize a wide array of walking tasks and sit-stand transitions, demonstrating the efficiency necessary for powered knee-ankle prostheses to become clinically viable. Clinical impact: This paper introduces a clinical tuning interface that simplifies the tuning process for multimodal robotic prosthetic legs, reducing the time required from several hours to just 20 minutes thus improving clinical feasibility.

关键词

KinematicsAnkleElectrical impedanceProsthesisComputer scienceKnee prosthesisImpedance controlPhysical medicine and rehabilitationOrthodonticsBiomedical engineering

相关论文

查看 LOCOMOTION 分类全部论文