Development of a Robotic Assembly for Analyzing the Instantaneous Axis of Rotation of the Foot Ankle Complex
K. N. Salb, D. M. Wido, Thomas E. Stewart, Denis J. DiAngelo
- 发表年份
- 2016
- 引用次数
- 9
- 访问权限
- 开放获取
摘要
Ankle instantaneous axis of rotation (IAR) measurements represent a more complete parameter for characterizing joint motion. However, few studies have implemented this measurement to study normal, injured, or pathological foot ankle biomechanics. A novel testing protocol was developed to simulate aspects of in vivo foot ankle mechanics during mid-stance gait in a human cadaveric specimen. A lower leg was mounted in a robotic testing platform with the tibia upright and foot flat on the baseplate. Axial tibia loads (ATLs) were controlled as a function of a vertical ground reaction force (vGRF) set at half body weight (356 N) and a 50% vGRF (178 N) Achilles tendon load. Two specimens were repetitively loaded over 10 degrees of dorsiflexion and 20 degrees of plantar flexion. Platform axes were controlled within 2 microns and 0.008 degrees resulting in ATL measurements within ±2 N of target conditions. Mean ATLs and IAR values were not significantly different between cycles of motion, but IAR values were significantly different between dorsiflexion and plantar flexion. A linear regression analysis showed no significant differences between slopes of plantar flexion paths. The customized robotic platform and advanced testing protocol produced repeatable and accurate measurements of the IAR, useful for assessing foot ankle biomechanics under different loading scenarios and foot conditions.
关键词
相关论文
Robot Motion Planning
Jean‐Claude Latombe
1991
Introduction to Robotics mechanics and Control
John Craig
1986
Closed-form solution of absolute orientation using unit quaternions
Berthold K. P. Horn
1987
Robot dynamics and control
Mark W. Spong
1989