首页 /研究 /Robotized ultrasound imaging of the peripheral arteries – a phantom study
LEARNING

Robotized ultrasound imaging of the peripheral arteries – a phantom study

Jannis Hagenah, Mark Kaschwich, Markus Kleemann, Verónica García‐Vázquez

发表年份
2020
引用次数
14

摘要

The first choice in diagnostic imaging for patients suffering from peripheral arterial disease (PAD) is 2D ultrasound (US). However, for a proper imaging process, a skilled and experienced sonographer is required. Additionally, it is a highly user-dependent operation. A robotized US system that autonomously scans the peripheral arteries has the potential to overcome these limitations. In this work, we extend a previously proposed system by a hierarchical image analysis pipeline based on convolutional neural networks (CNNs) in order to control the robot. The system was evaluated by checking its feasibility to keep the vessel lumen of a leg phantom within the US image while scanning along the artery. In 100% of the images acquired during the scan process the whole vessel lumen was visible. While defining an insensitivity margin of 2.74 mm, the mean absolute distance between vessel center and the horizontal image center line was 2.47 mm and 3.90 mm for an easy and complex scenario, respectively. In conclusion, this system presents the basis for fully automatized peripheral artery imaging in humans using a radiation-free approach.

关键词

Imaging phantomSonographerConvolutional neural networkLumen (anatomy)PeripheralComputer scienceUltrasoundBiomedical engineeringArtificial intelligenceComputer vision