首页 /研究 /High-Pressure Gel Loader for Capillary Array Electrophoresis Microchannel Plates
OTHER

High-Pressure Gel Loader for Capillary Array Electrophoresis Microchannel Plates

Jessica Scherer, Brian M. Paegel, Gary J. Wedemayer, Charles A. Emrich, Josephine T.-M. Lo, Igor L. Medintz, Richard A. Mathies

发表年份
2001
引用次数
32
访问权限
开放获取

摘要

Microfabricated capillary array electrophoresis (microCAE) microchannel plates are the next generation of bioanalytical separation devices. To fully exploit the capabilities of microCAE devices, supporting technology such as robotic sample loading, gel loading, microplate washing, and data analysis must be developed. Here, we describe a device for loading gel into radial capillary array electrophoresis microplates and for plate washing and drying. The microplates are locked into a loading module, and high-pressure helium is used to drive aqueous separation media or wash solutions into the microchannels through fixtures connected to the central anode reservoir. Microplates are rapidly (30 s to 5 min) loaded with separation media, such as 3%-4.8% linear polyacrylamide or 0.7%-3.0% hydroxyethyl cellulose, for electrophoresis. The effective and rapid gel-filling and plate-cleaning methods together with short electrophoretic analysis times (2-30 min) make microCAE systems versatile and powerful nucleic acid analysis platforms.

关键词

MicrochannelCapillary electrophoresisChromatographyElectrophoresisHydroxyethyl celluloseMaterials scienceCapillary actionMicrofluidicsFree-flow electrophoresisLoader

相关论文

查看 OTHER 分类全部论文