Fully automated robotic method for the screening of polychlorinated biphenyls in used mineral oils
A. Velasco‐Arjona, M. D. Luque de Castro, A. Izquierdo
- Year
- 1998
- Citations
- 7
Abstract
A screening method for the determination of polychlorinated biphenyls (PCBs) in waste mineral oils based on dechlorination by sodium and potentiometric measurement is proposed. The method is fully automated by a robotic station in which the robot performs the weighing of the sample, adds the reagents, develops a liquid–liquid extraction and obtains the chloride concentrations using a chloride-selective electrode. The data are acquired and treated by the computer. The method affords a detection limit of 1 µg g–1 chloride and a precision, expressed as relative standard deviation, of 3.2% which corresponds to a detection limit of 12 µg g–1 for Aroclor 1242. This detection limit is far below the regulatory limit of 50 µg g–1 which establishes whether or not a waste oil is contaminated by PCBs. The sample throughput is 12 h–1.