Home /Research /Fully automated robotic method for the screening of polychlorinated biphenyls in used mineral oils
OTHER

Fully automated robotic method for the screening of polychlorinated biphenyls in used mineral oils

A. Velasco‐Arjona, M. D. Luque de Castro, A. Izquierdo

Year
1998
Citations
7

Abstract

A screening method for the determination of polychlorinated biphenyls (PCBs) in waste mineral oils based on dechlorination by sodium and potentiometric measurement is proposed. The method is fully automated by a robotic station in which the robot performs the weighing of the sample, adds the reagents, develops a liquid–liquid extraction and obtains the chloride concentrations using a chloride-selective electrode. The data are acquired and treated by the computer. The method affords a detection limit of 1 µg g–1 chloride and a precision, expressed as relative standard deviation, of 3.2% which corresponds to a detection limit of 12 µg g–1 for Aroclor 1242. This detection limit is far below the regulatory limit of 50 µg g–1 which establishes whether or not a waste oil is contaminated by PCBs. The sample throughput is 12 h–1.

Keywords

Detection limitMineral oilChemistryReagentRelative standard deviationExtraction (chemistry)ChromatographyChlorideOrganic chemistry

Related papers

Browse all OTHER papers