Accuracy in planar cutting of bones: an ISO‐based evaluation
Olivier Cartiaux, Pierre‐Louis Docquier, Bernard Francq, Benoît Raucent, Etienne Dombre, Xavier Banse
- Year
- 2009
- Citations
- 26
- Access
- Open access
Abstract
BACKGROUND: Computer- and robot-assisted technologies are capable of improving the accuracy of planar cutting in orthopaedic surgery. This study is a first step toward formulating and validating a new evaluation methodology for planar bone cutting, based on the standards from the International Organization for Standardization. METHODS: Our experimental test bed consisted of a purely geometrical model of the cutting process around a simulated bone. Cuts were performed at three levels of surgical assistance: unassisted, computer-assisted and robot-assisted. We measured three parameters of the standard ISO1101:2004: flatness, parallelism and location of the cut plane. RESULTS: The location was the most relevant parameter for assessing cutting errors. The three levels of assistance were easily distinguished using the location parameter. CONCLUSIONS: Our ISO methodology employs the location to obtain all information about translational and rotational cutting errors. Location may be used on any osseous structure to compare the performance of existing assistance technologies.