首页 /研究 /Turbidimetric determination of orthophosphate in digested plant material by flow-injection analysis
OTHER

Turbidimetric determination of orthophosphate in digested plant material by flow-injection analysis

Maria-Celeste T. Diniz, André Fernando de Oliveira, Orlando Fatibello‐Filho

发表年份
2000
引用次数
2

摘要

A simple, fast, and low-cost flow-injection system is proposed for the determination of orthophosphate in digested plant material. The determination is based on the precipitation reaction of orthophosphate with 0.05 mol L−1 zinc solution in 0.05 mol L−1 triethanolamine buffer (TEA/TEAH+, pH 6.0) solution containing 0.025% (w/v) poly-(vinylalcohol). The zinc orthophosphate suspension produced was directly measured at 410 nm. Under the best experimental conditions, orthophosphate was determinated in the concentration range from 5 to 60 mg P L−1, and the relative standard deviations were less than 1.6% for 10 and 20 mg P L−1 solutions (n = 9). The analytical frequency was 180 h−1. The results obtained using the proposed FIA procedure and those obtained using a standard spectrophotometric procedure agree at the 95% confidence level. © 2000 John Wiley & Sons, Inc. Lab Robotics and Automation 12:236–240, 2000

关键词

TriethanolamineFlow injection analysisRelative standard deviationChemistryZincPrecipitationChromatographyAnalytical Chemistry (journal)Buffer solutionSuspension (topology)