首页 /研究 /Preliminary study of a new macro‐micro robot system for dental implant surgery: Design, development and control
SURGICAL

Preliminary study of a new macro‐micro robot system for dental implant surgery: Design, development and control

Yan Wang, Wei Wang, Yueri Cai, Qiming Zhao, Yuyang Wang

发表年份
2023
引用次数
3
访问权限
开放获取

摘要

BACKGROUND: This study aims to develop a new dental implant robotic system (DIRS) to relieve the burden and enhance the quality of dental implant surgery. METHODS: The implanting actuator and system controller are two parts. The implanting actuator is designed on the basis of the RCM mechanism, with its kinematics modelled. Besides, a multi-DOF admittance control strategy and a hybrid position-admittance control strategy were designed, endowing the actuator with environmental compliance. RESULTS: In force sensing, about 97% of mixed force/torque are eliminated. Then, 30 groups of implantation are done, of which 15 groups are simple implantation, while another 15 groups are force-based implantation. The results show that the average contact force/torque can be reduced by 73.03% and 62.66%, and the peak contact force/torque can be reduced by 68.26% and 50.46%. CONCLUSIONS: The results of preliminary experiments validate the effectiveness of DIRS, which has great potential to assist dentists with higher efficiency, better quality, and lower burden.

关键词

ActuatorKinematicsAdmittanceTorqueController (irrigation)ImplantComputer scienceMacroRobotSimulation

相关论文

查看 SURGICAL 分类全部论文