Study on the implementation of flow titrations in sequential injection analysis

José Manuel Estela, F. Más, A. Cladera, V. Cerd�

发表年份
1999
引用次数
4

摘要

Several variables that affect flow spectrophotometric titrations in sequential injection analysis (SIA) systems were studied by using titration of a strong acid with a strong base as a model; blue bromothymol was used as an indicator and distilled water was used as a carrier. Both possibilities of performing titrations on SIA systems according to the aspiration sequences (a) titrant B/sample/titrant A and (b) titrant/sample are considered. In case (a) the possibilities regarding the use of different volumes and/or different concentrations of titrant reagents were studied and are discussed. The established methodology allows hydrochloric acid determinations with 0.1 M sodium hydroxide within the range 32–1000 mM by means of logarithmic calibration curves. The analysis rate is of 5 min/sample. © 1999 John Wiley & Sons, Inc. Lab Robotics and Automation 11: 207–216, 1999