A Pneumatic Force Sensor for Enhanced Force Feedback in Robotic Colonoscopy

Korn Borvorntanajanya, Jialei Shi, Jabed F Ahmed, Enrico Franco, Ferdinando Rodriguez y Baena

发表年份
2025
引用次数
4

摘要

Colonoscopy is often associated with significant discomfort, which can be mitigated by incorporating haptic feedback into robotic systems. This paper presents the design and experimental evaluation of a force sensing system integrated with haptic feedback for robotic colonoscopy. The system utilizes pneumatic sensing pouches mounted on the tip of a robotic colonoscope to capture contact forces in real-time and provide feedback to the operator via a haptic joystick. The pouch sensor demonstrates accuracy, with a maximum RMS error of 0.2 N during rigid phantom testing, while achieving 85% phase classification accuracy in detecting on-load and off-load states. Haptic force feedback significantly reduces peak interaction forces, lowering the maximum contact force from 1 N to 0.6 N without compromising task duration. Experiments in a soft colon phantom confirm the system’s ability to detect forces in anatomically representative environments, demonstrating its potential for clinical use.