Towards synergistic control of hands-on needle insertion with automated needle steering for MRI-guided prostate interventions
Marek Wartenberg, Niravkumar Patel, Gang Li, Gregory S. Fischer
- 发表年份
- 2016
- 引用次数
- 5
摘要
A significant hurdle of accurate needle tip placement in percutaneous needle-based prostate interventions is unmodeled needle deflection and tissue deformation during insertion. This paper introduces a robotic platform for developing synergistic, cooperatively controlled needle insertion algorithms decoupled from closed-loop image-guided needle steering. Shared control of the surgical workspace through human-robot synergy creates a balance between the accuracy of robotic autonomy while still providing ultimate control of the procedure to the physician. Validation tests were performed using camera-based image-guided feedback control of needle steering with cooperative hands-on needle insertion. Locations were targeted inside a transparent gelatin phantom with an average total error of 2.68 ± 0.34mm and in-plane error of 2.59 ± 0.30mm.