Small bowel obstruction caused by self-anchoring suture used for peritoneal closure following robotic inguinal hernia repair
Faraz A. Khan, Asra Hashmi, David A. Edelman
- 发表年份
- 2016
- 引用次数
- 9
- 访问权限
- 开放获取
摘要
Laparoscopic inguinal herniorraphy is a commonly performed procedure given the reported decrease in pain and earlier return to activity when compared with the open approach. Moreover, robotic assistance offers the operating surgeon considerable ergonomic advantages, making it an attractive alternative to conventional laparoscopic herniorraphy. Robotic herniorraphy utilizes the transabdominal preperitoneal approach where following repair peritoneal closure is necessary to avoid mesh exposure to the viscera. Self-anchoring sutures are frequently used to this end given the ease of use and knotless application. We present an unusual case of post-operative small bowel obstruction following robotic inguinal hernia repair caused by the self-anchoring suture used for peritoneal closure. This patient presented 3 days post-procedure with symptoms and cross-sectional imaging indicative of small bowel obstruction with a clear transition point. Underwent laparoscopic lysis of a single adhesive band originating from the loose intraperitoneal end of the suture leading to resolution of symptoms.