A quantitative accuracy assessment of the use of a rigid robotic arm in navigated placement of 726 pedicle screws

Carlo Alberto Benech, Rosa Perez, Franco Benech, Torrey Shirk, Brandon S. Bucklen

发表年份
2022
引用次数
9
访问权限
开放获取

摘要

BACKGROUND: Traditional minimally invasive fluoroscopy-based techniques for pedicle screw placement utilize guidance, which may require fluoroscopic shots. Computerized tomography (CT) navigation results in more accurate screw placement. Robotic surgery seeks to establish access and trajectory with greater accuracy. OBJECTIVE: This study evaluated the screw placement accuracy of a robotic platform. METHODS: Demographic data, preoperative/postoperative CT scans, and complication rates of 127 patients who underwent lumbosacral pedicle screw placement with minimally invasive navigated robotic guidance using preoperative CT were analyzed. RESULTS: On the GRS scale, 97.9% (711/726) of screws were graded A or B, 1.7% (12/726) of screws graded C, 0.4% (3/726) of screws graded D, and 0% graded E. Average offset from preoperative plan to final screw placement was 1.9 ± 1.5 mm from tip, 2.2 ± 1.4 mm from tail and 2.9 ± 2.3° of angulation. CONCLUSIONS: Robotic-assisted surgery utilizing preoperative CT workflow with intraoperative fluoroscopy-based registration improves pedicle screw placement accuracy within a patient's pedicles.

关键词

MedicineFluoroscopyRobotic surgeryLumbosacral jointSurgeryComputed tomographyRadiology