A Novel Method for Adjusting the Taper and Adaption of Automatic Tooth Preparations with a High-Power Femtosecond Laser
Fusong Yuan, Shanshan Liang, Peijun Lyu
- 发表年份
- 2021
- 引用次数
- 9
- 访问权限
- 开放获取
摘要
This study explored the effect of the light-off delay setting in a robotically controlled femtosecond laser on the taper and adaption of resin tooth preparations. Thirty resin teeth (divided into six equal groups) were studied under different light-off delay conditions. Tapers from six vertical sections of the teeth were measured and compared among the light-off delay groups. The mean taper decreased from 39.268° ± 4.530° to 25.393° ± 5.496° as the light-off delay increased (p < 0.05). The average distance between the occlusal surfaces of the scanned data and the predesigned preparation data decreased from 0.089 ± 0.005 to 0.013 ± 0.030 μm as the light-off delay increased (p < 0.05). The light-off delay of the femtosecond laser is correlated with the taper and adaption of automatic tooth preparations; this setting needs to be considered during automatic tooth preparation.