Analysis of intrafraction motion in CyberKnife-based stereotaxy using mask based immobilization and 6D-skull tracking<sup>†</sup>.

Tejinder Kataria, Kushal Narang, Deepak Gupta, Shyam Singh Bisht, Ashu Abhishek, Shikha Goyal, Trinanjan Basu, K.P. Karrthick

发表年份
2016
引用次数
12

摘要

Purpose: Analysis of intrafraction motion in patients with intracranial targets treated with frameless, mask based stereotactic radiosurgery / radiotherapy using standard couch and 6D-skull tracking on CyberKnife. Materials and methods: Twenty-seven treatment datasets of fifteen patients were analyzed. For each sequential pair of images, the correction to the target position (position "offset") in six-degrees of motion was obtained. These offsets were used to calculate intrafraction shifts, and their statistical distribution. PTV margins were calculated, based on Van Herk formula. Results: The mean ± 1 SD intrafraction translationals were 0.27±0.61mm in left-right, 0.24±0.62mm in antero-posterior and 0.14±0.24mm in supero-inferior direction, and rotations were 0.13±0.21 degrees roll, 0.18±0.25 degrees pitch and 0.28±0.44 degrees yaw. Most intrafraction shifts were ≤ 1mm and 1 degree. Fourteen instances of intrafraction shifts exceeding the robotic correction threshold were noted. Calculated PTV margins were 1mm, 1mm and 0.4mm for for left-right, antero-posterior and supero-inferior directions, respectively. Conclusions: CyberKnife 6D-skull tracking with 1mm PTV margin effectively compensates for intrafraction motion. The occasional large intrafraction movements may assume significance for techniques not employing intrafraction motion management.

关键词

CyberknifeRadiosurgeryNuclear medicineMedicineRadiation therapyRadiology

相关论文

查看 SURGICAL 分类全部论文