Robotic chromatography: development and evaluation of automated instrumentation for assay of glycohemoglobin

Christopher D. Herold, Kirk Andree, David A. Herold, Robin A. Felder

发表年份
1993
引用次数
14

摘要

The measurement of glycohemoglobin (GHb) by boronate affinity chromatography is useful in monitoring long-term glucose control in diabetic subjects. The inherent disadvantage of this method is the hands-on time required because the hemoglobin fractions are separated on individual disposable columns. To overcome this disadvantage, we have programmed a Hamilton Microlab 2200 automated pipetting cartesian robot to complete the procedure, from the aspiration of blood from the sample-collection tube to the transfer of the separated hemoglobin fractions to a microtiter plate for absorbance measurement. This automated robotic system can analyze 96 specimens, including patients' samples and control material, in approximately 3 h. The precision (CV) of the method ranged from 1.6% to 3.5% within-run and from 2.7% to 3.5% day-to-day. The results correlated with those obtained with the Accuflex semiautomated robot, which used the identical disposable column, and those obtained with a Primus high-performance liquid chromatograph, which used a regenerated microparticle column. Automation of the GHb procedure allowed improved throughput, reduced labor cost, improved precision, and offered greater laboratory safety.