首页 /研究 /Design and Experimental Setup of a Robotic Medical Instrument for Brachytherapy in Non-Resectable Liver Tumors
SURGICAL

Design and Experimental Setup of a Robotic Medical Instrument for Brachytherapy in Non-Resectable Liver Tumors

Paul Tucan, Călin Vaida, Daniel Horvath, Andrei Caprariu, Alin Burz, Bogdan Gherman, Stefan Iakab, Doina Pîslă

发表年份
2022
引用次数
30
访问权限
开放获取

摘要

This paper presents a study regarding the design and the experimental setup of a medical robotic system for brachytherapy using tribology analysis. The robotic system is composed of a collaborative robotic arm and a multi-needle brachytherapy instrument controlled using a unified control system embedding a haptic device and force-feedback. This work is oriented towards identifying the technical characteristics of the system components to determine the accuracy of the procedure, as well as using different scenarios for needle insertion in ex vivo porcine liver tissue in order to determine the forces required for insertion and extraction of the needle and the friction coefficient that accompanies the previously mentioned forces. Subsequent to the computation of the friction forces, the normal forces and the wear during the needle insertion are determined with the scope of predicting the lifecycle of some components of the medical device.

关键词

BrachytherapyHaptic technologyScope (computer science)Robotic armComputer scienceBiomedical engineeringRobotComputationSimulationMedical physics