High three‐dimensional accuracy of component placement and lower limb alignment using a robotic arm‐assisted system and gap‐balancing instrument in total knee arthroplasty
Emanuele Diquattro, Robert Prill, Mikhail Salzmann, Francesco Traina, Roland Becker
- 发表年份
- 2024
- 引用次数
- 33
摘要
PURPOSE: It was hypothesized that robotic arm-assisted total knee arthroplasty (RA-TKA) using additionally a gap-balancing instrumentation will show high accuracy in executing the planning in femoral and tibial component placement throughout the range of knee motion (ROM) during TKA surgery. METHODS: Prospectively collected data were analysed for patients undergoing RA-TKA. A cruciate retaining cemented design was implanted using the MAKO® robotic system. Lower limb alignment at 0°, 30°, 45°, 60° and 90° of flexion was recorded at the beginning of surgery and finally after implantation of the components. A ligament tensioner was inserted after tibial precut to measure the extension and flexion gap, and final component placement was planned based on 3D CT images. Femoral and tibial component placement was measured in all three planes. RESULTS: A total of 104 patients were included (mean age 69.4 ± 9 years; 44 male, 60 female). The difference in component placement after planning and final implantation showed less valgus of 0.7° ± 1.4° (p < 0.001), less external rotation of 0.6° ± 1.9° (p = 0.001) and less flexion of 0.9° ± 1.8° (p < 0.001) for the femoral component. The tibial component was placed in more varus of 0.2° ± 0.9° (p = 0.056) and more posterior slope of 0.5° ± 0.9° (p < 0.001). The lower limb alignment in extension was 4.4° ± 5.2° of varus of the native knee and changed to 1.2° ± 1.9° of varus after TKA (p < 0.01). CONCLUSION: Robotic-assisted TKA helps to achieve the target of alignment and component placement very close to the planning. It allows optimal component placement of off-the-shelf implants respecting patient's specific anatomy. LEVEL OF EVIDENCE: Level II.
关键词
相关论文
Robots and Jobs: Evidence from US Labor Markets
Daron Acemoğlu, Pascual Restrepo
2019
Reach and grasp by people with tetraplegia using a neurally controlled robotic arm
Leigh R. Hochberg, Daniel Bacher, Beata Jarosiewicz 等 11 位作者
2012
Campbell-Walsh urology
Alan J. Wein editor-in-chief
2012
Stroke rehabilitation
Peter Langhorne, Julie Bernhardt, Gert Kwakkel
2011