首页 /研究 /Quantifying workspace and forces of surgical dissection during robot‐assisted neurosurgery
SURGICAL

Quantifying workspace and forces of surgical dissection during robot‐assisted neurosurgery

Yaser Maddahi, Liu Shi Gan, Kourosh Zareinia, Sanju Lama, Nariman Sepehri, Garnette R. Sutherland

发表年份
2015
引用次数
38
访问权限
开放获取

摘要

BACKGROUND: A prerequisite for successful robot-assisted neurosurgery is to use a hand-controller matched with characteristics of real robotic microsurgery. This study reports quantified data pertaining to the required workspace and exerted forces of surgical tools during robot-assisted microsurgery. METHODS: A surgeon conducted four operations in which the neuroArm surgical system, an image-guided computer-assisted manipulator specifically designed to perform robot-assisted neurosurgery, was employed to surgically remove brain tumors. The position, orientation, and exerted force of surgical tools were measured during operations. RESULTS: Workspace of the neuroArm manipulators, for the cases studied, was 60×60×60 mm(3) while it offered orientation ranges of 103°, 62° and 112°. The surgical tools exerted a maximum force of 1.86 N with frequency band of less than 20 Hz. CONCLUSIONS: This data provides important information specific to neurosurgery that can be used to select among commercially available, or further design a customized, haptic hand-controller for robot-assisted neurosurgical systems. Copyright © 2015 John Wiley & Sons, Ltd.

关键词

WorkspaceMicrosurgeryNeurosurgeryOrientation (vector space)RobotHaptic technologyComputer scienceController (irrigation)Medical physicsSurgery

相关论文

查看 SURGICAL 分类全部论文