Lateral temperature spread of monopolar, bipolar and ultrasonic instruments for robot‐assisted laparoscopic surgery
Lukas J. Hefermehl, Remo A. Largo, Thomas Hermanns, Cédric Poyet, Tullio Sulser, Daniel Eberli
- 发表年份
- 2013
- 引用次数
- 99
摘要
OBJECTIVE: To assess critical heat spread of cautery instruments used in robot-assisted laparoscopic (RAL) surgery. MATERIALS AND METHODS: Thermal spread along bovine musculofascial tissues was examined by infrared camera, histology and enzyme assay. Currently used monopolar, bipolar and ultrasonic laparoscopic instruments were investigated at various power settings and application times. The efficacy of using an additional Maryland clamp as a heat sink was evaluated. A temperature of 45 °C was considered the threshold temperature for possible nerve damage. RESULTS: Monopolar instruments exhibited a mean (sem) critical thermal spread of 3.5 (2.3) mm when applied at 60 W for 1 s. After 2 s, the spread was >20 mm. For adjustable bipolar instruments the mean (sem) critical thermal spread was 2.2 (0.6) mm at 60 W and 1 s, and 3.6 (1.3) mm at 2 s. The PK and LigaSure forceps had mean (sem) critical thermal spreads of 3.9 (0.8) and 2.8 (0.6) mm respectively, whereas the ultrasonic instrument reached 2.9 (0.8) mm. Application of an additional Maryland clamp as a heat sink, significantly reduced the thermal spread. Histomorphometric analyses and enzyme assay supported these findings. CONCLUSIONS: All coagulation devices used in RAL surgery have distinct thermal spreads depending on power setting and application time. Cautery may be of concern due to lateral temperature spread, causing potential damage to sensitive structures including nerves. Our results provide surgeons with a resource for educated decision-making when using coagulation devices during robotic procedures.
关键词
相关论文
Robots and Jobs: Evidence from US Labor Markets
Daron Acemoğlu, Pascual Restrepo
2019
Skin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes
Darren J. Lipomi, Michael Vosgueritchian, Benjamin C. K. Tee 等 7 位作者
2011
Stretchable, Skin‐Mountable, and Wearable Strain Sensors and Their Potential Applications: A Review
Morteza Amjadi, Ki‐Uk Kyung, Inkyu Park 等 4 位作者
2016
Reach and grasp by people with tetraplegia using a neurally controlled robotic arm
Leigh R. Hochberg, Daniel Bacher, Beata Jarosiewicz 等 11 位作者
2012