Design of an endoluminal NOTES robotic system
Daniel J. Abbott, C. Becke, Richard Rothstein, William J. Peine
- 发表年份
- 2007
- 引用次数
- 194
摘要
Natural orifice transluminal endoscopic surgery, or NOTES, allows for exceedingly minimally invasive surgery but has high requirements for the dexterity and force capabilities of the tools. An overview of the ViaCath System is presented. This system is a first generation teleoperated robot for endoluminal surgery and consists of a master console with haptic interfaces, slave drive mechanisms, and 6 degree-of-freedom, long-shafted flexible instruments that run alongside a standard gastroscope or colonoscope. The system was validated through animal studies. It was discovered that the devices were difficult to introduce into the GI tract and manipulation forces were insufficient. The design of a second generation system is outlined with improvements to the instrument articulation section and a steerable overtube. Results of basic evaluation tests performed on the tools are also presented.
关键词
相关论文
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
A new optimizer using particle swarm theory
R.C. Eberhart, James Kennedy
2002